Electronic Scale Integration#

The Weigh and Dispense pages make some assumptions about the communication with the electronic scale through the application message bus. This page describes the structure of the messages and the integration workflows for positive and negative weighing.

Info

The scale Resource must integrate with the system, so that the Weigh and Dispense wizard receives the values of the scale automatically. For the configuration of the scale, see Configure Weigh and Dispense.

Message Structures#

1. Instrument Configuration#

Type Subject
RequestReply Measurement.[ResourceId].Setup

Table: Instrument configuration message

{
  "SendToSubject": "CMF.Resource.[ResourceId].[Guid]",
  "SetTare": true
}

The SetTare property is true or false.

2. Begin Weighing#

Type Subject
RequestReply Measurement.[ResourceId].Start

Table: Begin weighing message

{
  "SendToSubject": "CMF.Resource.[ResourceId].[Guid]",
  "Timeout": 6000,
  "ResetTare": true
}

The ResetTare property is true or false.

3. Stop Weighing#

Type Subject
RequestReply Measurement.[ResourceId].Stop

Table: Stop weighing message

{
  "SendToSubject": "CMF.Resource.[ResourceId].[Guid]"
}

4. Notify Weight Value#

Type Subject
RequestReply The SendToSubject that the Begin Weighing request defines

Table: Notify weight value message

{
  "Timestamp": "2019-02-05T11:17:51.626Z",
  "Value": 1234.5678,
  "IsStable": false,
  "Extra": {
    "TareValue": 0,
    "WeightIncludesTare": false,
    "WeighingUnits": "Kg"
  }
}

5. Interface Events#

Type Subject
RequestReply Measurement.[ResourceId].Event

Table: Interface events message

The interface sends this message structure for the following events:

  • Dispense and Continue
  • Dispense and Reset
  • Dispense
  • Begin weighing
  • Complete weighing
  • Scale change
  • Weighing mode change, from positive to negative or from negative to positive
  • Undo
  • Page close, when the operator closes the page abruptly
  • Ingredient selection, when the operator switches ingredients or opens the Weigh and Dispense page

The following examples show the message for three events.

Example 1: Weighing mode change.

{
  "OperationType": 10,
  "Data": {
    "WeighingMode": "Negative"
  }
}

Example 2: Ingredient selection.

{
  "OperationType": 13,
  "Data": {
    "tareValue": 0,
    "dispensedQuantity": 0,
    "requiredQuantity": 3,
    "bomProductName": "BomAutomationWD#IG.4ToFlour.6",
    "bomProductId": "2002110227330000006"
  }
}

Example 3: Generic event.

{
  "OperationType": 6
}

Integration Workflows#

The following workflows describe the interaction between the Weigh and Dispense GUI, the equipment integration, and the scale.

Positive Weighing#

Sequence diagram of the integration workflow for positive weighing between the GUI, the equipment integration, and the scale.

  1. When the Weigh and Dispense GUI starts, it sends a Start request to the equipment integration and sets Reset to true. The equipment integration resets the tare, if the scale supports it, and stores the TareValue. It reports the weight and the tare to the GUI with ReportValue, including whether the weight includes the tare. The GUI sets GUIWeightValue and GUITareValue.
  2. When the operator selects Set Tare & Begin, the GUI sends Configure (Positive). If the scale supports it, the equipment integration performs a remote tare and stores the TareValue. The GUI sets GUITareValue to TareValue.
  3. The equipment integration reads the weight of the scale with GetValue and reports it to the GUI. When the weight includes the tare, the GUI calculates GUIWeigh as WeightValue minus TareValue. Otherwise, GUIWeigh is WeightValue.
  4. On a partial or complete dispense, the GUI sends the Start request again, sets Reset to true, and the workflow repeats.

Negative Weighing#

Sequence diagram of the integration workflow for negative weighing between the GUI, the equipment integration, and the scale.

  1. When the Weigh and Dispense GUI starts, it sends an Initialize request to the equipment integration and sets Reset to true. The equipment integration resets the tare, if the scale supports it, and stores the TareValue. It reports the weight and the tare to the GUI with ReportValue. The GUI sets GUIWeightValue and GUITareValue.
  2. When the operator begins weighing, the GUI performs the action alone and sends no message to the equipment integration. When the weight includes the tare, the GUI sets GUIGrossWeigh to WeightValue. Otherwise, it sets GUIGrossWeigh to WeightValue plus TareValue.
  3. For each new ReportValue, the GUI calculates GUIWeightValue. When the weight includes the tare, GUIWeightValue is GUIGrossWeigh minus WeightValue. Otherwise, it is GUIGrossWeigh minus the sum of WeightValue and TareValue.
  4. On a partial or complete dispense, the GUI sends the Initialize request again, sets Reset to true, and the workflow repeats.

Expected Result#

You know the messages and the workflows that the equipment integration must support, so that the Weigh and Dispense pages receive the values of an electronic scale.